Comparison of commercially available serologic kits for the detection of celiac disease
Afzal J Naiyer1, Lincoln Hernandez, Edward J Ciaccio
1Department of Medicine and Pathology, Columbia University, New York, NY 10032, USA.
Background:
The sensitivity and specificity of current antihuman tissue transglutaminase (tTG) IgA assays used to detect celiac disease reportedly approach 100%. In addition, the sensitivity of new generation deamidated gliadin peptide (alpha-DGP) antibody assays has also been reported to be similar to the tTG IgA assays. In routine clinical practice, however, the sensitivities and specificities of these tests for diagnosing celiac disease seem to be lower.
Aim:
We analyzed sensitivities and specificities of 4 IgA tTG and 3 deamidated gliadin peptide (alpha-DGP) kits.
Methods:
The performance of 4 tTG IgA assays, A: Inova (Hu red blood cell), B: Binding site (rHu Ag), C: Eurospital (rHu Ag), D: Immco (rHu Ag) and 3 Inova alpha-DGP assays, E: alpha-DGP-IgA, F: alpha-DGP-IgG, and G: alpha-DGP-IgA+G was evaluated using sera from different subsets of celiac disease patients and controls; group 1: active celiac disease n=28, group 2: gluten-free diet n=54, group 3: healthy controls n=40, group 4: disease controls n=57(Crohn's disease n=17, chronic hepatitis n=40).
Results:
Using the manufacturer's cut-off values, the sensitivities and specificities of different kits ranged from 71.4% to 96.4% and 87.5% to 100%, respectively. When group 1 was compared with disease controls, sensitivities remained the same but specificities decreased. Receiver operating characteristic plot derived cut-off values modified decision thresholds in all assays except kit (G). Kappa analysis demonstrated variable degrees of agreement. All assays demonstrated higher sensitivities for patients with higher grades of villous atrophy.
Conclusions:
Overall sensitivity was at or below 90%, which is lower than that reported in the literature. Performance of the recombinant and red blood cell antigen-based tTG assays was similar, whereas the alpha-DGP assays demonstrated lower values. Receiver operating characteristic plot derived cut-off values altered test results. Many factors affect the results of these tests and clinicians should be aware of their limitations.
More Related Videos
06:50A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
09:12A Protein Microarray Assay for Serological Determination of Antigen-specific Antibody Responses Following Clostridium difficile Infection
Published on: June 15, 2018
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Serum Laboratory Studies, Stool Test, Breath Test