[Interference of labile glycated hemoglobin on HbA(1c) assay by a high pressure liquid chromatography method]

J Szymezak1, N Leroy, E Lavalard

  • 1Laboratoire de biochimie et de recherche pédiatriques, American Memorial Hospital, CHU de Reims.

HbA(1c) assay by high pressure liquid chromatography remains submitted to interferences, among which that of labile HbA(1c) in 1 to 2% of samples. We have evaluated the interference of labile HbA(1c) on HbA(1c) assay using Variant II analyzer (Biorad), by in vitro formation of labile glycated haemoglobin and by evaluation of two protocols of elimination of labile HbA(1c) (wash and incubation of red blood cells in saline solution, or incubation in the wash/dilution solution of the analyzer). Levels of labile HbA(1c) higher than 4.5 % lead to underestimation of HbA(1c). The different protocols tested proved efficient and were adapted to routine conditions. The fastest method is the incubation of red blood cells in the wash/dilution solution for at least two hours, or more if labile fraction is unusually high.

Related Concept Videos

High-Performance Liquid Chromatography: Introduction01:11

High-Performance Liquid Chromatography: Introduction

High-performance liquid chromatography(HPLC), formerly referred to as High-pressure liquid chromatography, is a powerful technique used to separate, identify, and quantify components in complex mixtures. The term "high pressure" refers to using high pressure to push the liquid mobile phase through the tightly packed columns.
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation00:57

High-Performance Liquid Chromatography: Instrumentation

High-performance liquid chromatography, or HPLC, is an analytical technique that separates liquid samples under high pressures. An HPLC instrument consists of glass bottles for storing solvents called mobile phase reservoirs. HPLC-grade solvents are used to maintain high purity, and the dissolved gases are removed using a degasser, such as a vacuum pumping system or sparging with helium. The solvents are then pumped into the analytical column using a screw-driven syringe or reciprocating pumps.
Hyperglycemia01:29

Hyperglycemia

Hyperglycemia is an abnormally high blood glucose level. It is diagnosed by fasting glucose ≥126 mg/dL, 2-hour oral glucose tolerance test (or OGTT) ≥200 mg/dL, random glucose ≥200 mg/dL with symptoms, or HbA1c ≥6.5%. However, HbA1c results may be unreliable in certain conditions, such as anemia or hemoglobinopathies, and the diagnosis should be confirmed unless classic symptoms are present. Postprandial hyperglycemia is typically considered significant when glucose levels exceed 180 mg/dL two...
High-Performance Liquid Chromatography: Elution Process01:05

High-Performance Liquid Chromatography: Elution Process

In High-Performance Liquid Chromatography (HPLC), the elution process is critical to the separation of analytes and the quality of chromatographic results. Elution describes how compounds move through the column and separate based on their interactions with the mobile and stationary phases. This process determines the resolution, peak shape, and retention times in the chromatogram, which are essential for identifying and quantifying components in complex mixtures. Understanding the elution...
Atomic Absorption Spectroscopy: Interference01:25

Atomic Absorption Spectroscopy: Interference

Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...