Image quality vs radiation dose of four cone beam computed tomography scanners
Objectives:
To evaluate image quality by examining segmentation accuracy and assess radiation dose for cone beam CT (CBCT) scanners.
Methods:
A skull phantom, scanned by a laser scanner, and a contrast phantom were used to evaluate segmentation accuracy. The contrast phantom consisted of a polymethyl methacrylate (PMMA) cylinder with cylindrical inserts of air, bone and PMMA. The phantoms were scanned on the (1) Accuitomo 3D, (2) MercuRay, (3) NewTom 3G, (4) i-CAT and (5) Sensation 16. The structures were segmented with an optimal threshold. Thicknesses of the bone of the mandible and the diameter of the cylinders in the contrast phantom were measured across lines at corresponding places in the CT image vs a ground truth. The accuracy was in the 95th percentile of the difference between corresponding measurements. The correlation between accuracy in skull and contrast phantom was calculated. The radiation dose was assessed by DPI(100,c) (dose profile integral (100,c)) at the central hole of a CT dose index (CTDI) phantom.
Results:
The results for the DPI(100,c) were 107 mGy mm for (1), 1569 mGy mm for (2), 446 mGy mm for (3), 249 mGy mm for (4) and 1090 mGy mm for (5). The segmentations in the contrast phantom were submillimeter accurate in all scanners. The segmentation accuracy of the mandible was 2.9 mm for (1), 4.2 mm for (2), 3.4 mm for (3), 1.0 mm for (4) and 1.2 mm for (5). The correlation between measurements in the contrast and skull phantom was below 0.37 mm.
Conclusions:
The best radiation dose vs image quality was found for the i-CAT.
More Related Videos
05:49Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT