Application of a Dual Upper Limb Task-Oriented Robotic System for the Functional Recovery of the Upper Limb in Stroke Patients
Published on: October 11, 2024
Poststroke upper-limb rehabilitation using 5 to 7 inserted microstimulators: implant procedure, safety, and efficacy
Ross Davis1, Owen Sparrow, Gregoire Cosendai
1Alfred Mann Foundation, Santa Clarita, CA 91355, USA. rosdavis@cfl.rr.com
Objective:
To investigate the feasibility of implanting microstimulators to deliver programmed nerve stimulation for sequenced muscle activation to recover arm-hand functions.
Design:
By using a minimally invasive procedure and local anesthesia, 5 to 7 microstimulators can be safely and comfortably implanted adjacent to targeted radial nerve branches in the arm and forearm of 7 subjects with poststroke paresis. The microstimulators' position should remain stable with no tissue infection and can be programmed to produce effective personalized functional muscle activity with no discomfort for a preliminary 12-week study. Clinical testing, before and after the study, is reported in the accompanying study.
Setting:
Microstimulator implantations in a sterile operating room.
Participants:
Seven adults, with poststroke hemiparesis of 12 months or more.
Intervention:
Under local anesthesia, a stimulating probe was inserted to identify radial nerve branches. Microstimulators were inserted by using an introducer and were retrievable for 6 days by attached suture. Each device was powered via a radiofrequency link from 2 external cuff coils connected to a control unit.
Main Outcome Measures:
To achieve low threshold values at the target sites with minimal implant discomfort. Microstimulators and external equipment were monitored over 12 weeks of exercise.
Results:
Seven subjects were implanted with 41 microstimulators, 5 to 7 per subject, taking 3.5 to 6 hours. Implantation pain levels were 20% more than anticipated. No infections or microstimulator failures occurred. Mean nerve thresholds ranged between 4.0 to 7.7 microcoulomb/cm(2)/phase over 90 days, indicating that cathodes were within 2 to 4 mm of target sites. In 1 subject, 2 additional microstimulators were inserted.
Conclusions:
Microstimulators were safely implanted with no infection or failure. The system was reliable and programmed effectively to perform exercises at home for functional restoration.