Mechanical properties of magnetic attachments for removable prostheses on teeth and implants
Arne F Boeckler1, Dean Morton, Carolin Ehring
1Centre for Dentistry and Oral Medicine, Department of Prosthodontics, Martin Luther University, Grosse Steinstrasse 19, Halle-Wittenberg, Germany. arne.boeckler@medizin.uni-halle.de
Purpose:
Magnetic attachments on teeth and implants may be used to improve stability, support, and retention of removable prostheses. Various forms of magnetic attachments are available, divided according to the design, the mechanical properties of the attachments, and the clinical indication. Recently developed attachment systems are small and promise improved retentive capacity, while existing magnetic attachments continue to be technologically modified and improved. This investigation reviewed and compared maximum retentive forces and characteristic curves for magnetic attachments indicated for use as root anchors and on implants.
Materials And Methods:
Twenty-four samarium-cobalt (SmCo) and neodym-iron-boron (NeFeB) magnetic attachments (12 tooth- and 12 implant-borne) were evaluated. Specimens were delivered by the manufacturers or fabricated according to their instructions. Five magnet pairs of each product and each combination were tested 10 times in a calibrated universal testing machine using a nonmagnetic test device (s = 40 mm, v = 20 mm/min). Results were recorded electronically and compared to manufacturers' details.
Results:
Maximum retentive forces for root keepers ranged from 1.4 to 6.6 N. Maximum retentive forces for magnetic attachments on implants ranged from 0.7 to 5.8 N. After a distance of 0.1 mm, a complete reversed distribution of the different systems became obvious. The retentive force provided by the manufacturer was achieved in one implant abutment, with retentive force (as compared to those provided by the manufacturers) for root keepers ranging between 42.5% and 92.9% and for implant abutments between 43.0% and 99.4%.
Conclusion:
There were differences between magnetic attachments for both the initial retentive capabilities and the characteristic curves. Recently introduced products provided relatively high initial retentive forces despite their small size. The measured retentive forces and the manufacturer's information differed in the majority of magnetic systems evaluated.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
07:32Measuring Maxillary Posterior Tooth Movement: A Model Assessment using Palatal and Dental Superimposition
Published on: February 23, 2024
Related Concept Videos
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...
Ferromagnetism
Magnetic Resonance Imaging