[Major influential factors of the micturition alert device dedicated to neurogenic bladders]
Jianhuo Wang1, Chunlin Hou, Xianyou Zheng
1Department of Orthopaedics, Changzheng Hospital, the Second Military Medical University, Shanghai, 200003, PR China.
Objective:
To study major influential factors of the micturition alert device dedicated to neurogenic bladders for the product design and clinical application of the device.
Methods:
One ferrite permanent magnet with thickness and diameter of 3 mm and 10 mm, respectively, and three NdFeB permanent magnets with the thickness of 3 mm and diameter of 10, 15 and 20 mm, respectively, were used. The effects of thickness of the abdominal wall as well as the position and type of permanent magnets on the micturition alert device dedicated to neurogenic bladders were measured in vitro simulated test, when the abdominal wall was set to 2, 3, 4, 5, 6, 7, 8 and 9 cm, respectively, and the position of permanent magnets was 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11 and 12 cm, respectively. The effect of the geomagnetic field on the device was measured under the condition that the thickness of the simulated abdominal wall was set to 2, 3, 4 and 5 cm, respectively, and the position of permanent magnets was 2, 3, 4, 5, 6, 7, 8, 9 and 10 cm, respectively.
Results:
The value showed in the warning unit was positively correlated with the position of the ferrite permanent magnet only when the thickness of the simulated abdominal wall was 2 cm (r=0.632, P < 0.05). The correlation between the value of the warning unit and the position of NdFeB permanent magnets was significant (r > 0.622, P < 0.05), which was intensified with the increasing diameter of NdFeB permanent magnets, but weakened with the increasing thickness of the simulated abdominal wall. The effect of the geomagnetic field was correlated with the exposition of the body, the position of the permanent magnet and the thickness of the abdominal wall.
Conclusion:
The major influential factors of the micturition alert device dedicated to neurogenic bladder include the magnetism and location of the permanent magnet, the thickness of the abdominal wall and the geomagnetic field. These factors are correlated with and affect each other. Reasonable allocation of these factors may optimize the device.
Related Concept Videos
The Micturition Reflex
The process begins with bladder filling, where the bladder wall stretches as urine accumulates. This stretching activates the urine storage reflex, mediated by the sacral spinal segments and the pontine storage center. Efferent sympathetic impulses stimulate the detrusor muscle to relax and the internal urethral sphincter to contract, facilitating urine...
Anatomy of the Genitourinary System II: Bladder and Urethra
Urinary Bladder
In males, the bladder is situated in front of the rectum, while in females, it is positioned anterior to the vagina and uterus. The bladder floor contains an inverted triangular area called the trigone, defined by the two ureteric...
Nursing Assessment of the Genitourinary System I: Health History
Urodynamic Studies: Uroflowmetry
Urinary Tract Infection IV: Nursing Management