Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Comparison of self-gated cine MRI retrospective cardiac synchronization algorithms
Grace M Nijm1, Alan V Sahakian, Steven Swiryn
1Department of Electrical Engineering and Computer Science, Northwestern University, Evanston, Illinois 60208, USA.
Purpose:
To determine whether improved self-gating (SG) algorithms can provide superior synchronization accuracy for retrospectively gated cine MRI.
Materials And Methods:
First difference, template matching, and polynomial fitting algorithms were implemented to improve the synchronization of MRI data using cardiac SG signals. Cine datasets were acquired during short-axis, two-, three-, and four-chamber cardiac MRI scans. The root-mean-square (RMS) error of SG synchronization positions compared to detected R-wave positions were calculated along with the mean square error (MSE) and peak signal-to-noise ratio (PSNR) comparing SG to electrocardiogram (ECG)-gated images. Overall image quality was also compared by two expert reviewers.
Results:
RMS errors were highest for the first difference method for all orientations. Improvements for both template matching and cubic polynomial fitting methods were significant for two-, three-, and four-chamber scans. MSE values were lower and PSNR were significantly higher for the cubic method compared to the first difference method for all orientations. Reviewers scored the images to be of comparable quality.
Conclusion:
Template matching and polynomial fitting improved the accuracy of cardiac cycle synchronization for two-, three-, and four-chamber scans; improvements in SG synchronization accuracy were reflected in improvements in analytical image quality. Implementation of robust postprocessing algorithms may bring SG approaches closer to clinical utilization.
More Related Videos
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Imaging Studies for Cardiovascular System IV: CMRI