Related Experiment Video
Updated: Jul 1, 2026

Efficient Nucleic Acid Extraction and 16S rRNA Gene Sequencing for Bacterial Community Characterization
Published on: April 14, 2016
Plasmid sonication improves sequencing efficiency and quality in the Beckman Coulter CEQ system
Mark Thompson1, Kelly Aukema, Dana O'Bryan
1College of Science and Management, University of Northern British Columbia, Prince George, British Columbia, Canada.
Abstract:
We report on an unexpectedly high rate of unreadable chromatograms from plasmid sequencing using Beckman Coulter's protocols, chemistry, and CEQ8000 instrument. Failed or poor quality plasmid sequence chromatograms were accompanied by a sharp drop, fluctuation, or steady decline in the current and a corresponding delay in signal counts beyond the time of capillary injection. We observe a correlation between the presence of supercoiled DNA and these sequencing problems. Herein we demonstrate that plasmid sonication, which is known to fragment supercoiled DNA, is an effective way to improve sequence phred20 read lengths to the point that they are not significantly different from Beckman Coulter's control template or enzymatically linearized plasmids.