Related Concept Videos

Computed Tomography01:10

Computed Tomography

Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
Imaging Studies III: Computed Tomography01:27

Imaging Studies III: Computed Tomography

DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...
Adjusting a Traverse01:12

Adjusting a Traverse

In the site survey of a four-sided traverse, internal angles are essential to ensure geometric accuracy. The survey revealed that the sum of the measured internal angles was 359 degrees and 48 minutes, which is 12 minutes less than the expected 360 degrees. This discrepancy signals an error likely arising from measurement inaccuracies during the fieldwork.To rectify this error, the adjustment process involved distributing the 12-minute shortfall equally across the four internal angles. By...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Chromatix: a differentiable, GPU-accelerated wave-optics library.

Nature methods·2026
Same author

Chromatix: a differentiable, GPU-accelerated wave-optics library.

bioRxiv : the preprint server for biology·2026
Same author

Racial differences in texture analysis features of the choriocapillaris.

Canadian journal of ophthalmology. Journal canadien d'ophtalmologie·2026
Same author

Observing biological spatio-angular structures and dynamics with statistical image reconstruction and polarized fluorescence microscopy.

bioRxiv : the preprint server for biology·2026
Same author

The origin of vertebrate teeth and evolution of sensory exoskeletons.

Nature·2025
Same author

A Bayesian Multivariate Spatial Point Pattern Model: Application to Oral Microbiome FISH Image Data.

ArXiv·2025

Related Experiment Video

Updated: Jul 1, 2026

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
07:13

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities

Published on: October 27, 2023

Correction for resolution nonuniformities caused by anode angulation in computed tomography.

Patrick J La Rivière1, Phillip Vargas

  • 1Department of Radiology, The University of Chicago, Chicago, IL 60637, USA. pjlarivi@midway.uchicago.edu

IEEE Transactions on Medical Imaging
|September 10, 2008
PubMed
Summary

This study introduces a method to improve computed tomography (CT) image resolution by correcting for focal spot angulation. This technique ensures more uniform image quality across the entire field-of-view.

More Related Videos

High-Accuracy Correction of 3D Chromatic Shifts in the Age of Super-Resolution Biological Imaging Using Chromagnon
08:18

High-Accuracy Correction of 3D Chromatic Shifts in the Age of Super-Resolution Biological Imaging Using Chromagnon

Published on: June 16, 2020

Related Experiment Videos

Last Updated: Jul 1, 2026

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
07:13

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities

Published on: October 27, 2023

High-Accuracy Correction of 3D Chromatic Shifts in the Age of Super-Resolution Biological Imaging Using Chromagnon
08:18

High-Accuracy Correction of 3D Chromatic Shifts in the Age of Super-Resolution Biological Imaging Using Chromagnon

Published on: June 16, 2020

Area of Science:

  • Medical Imaging
  • Radiological Physics
  • Computed Tomography

Background:

  • X-ray tubes use rotating anodes, beveled at shallow angles to manage heat and focal spot size.
  • Anode angulation causes variations in effective focal spot size across detector arrays, degrading peripheral image resolution.
  • This focal spot size variation leads to anisotropic resolution in fan-beam CT reconstructions.

Purpose of the Study:

  • To develop and describe a method for correcting focal spot angulation effects in fan-beam CT.
  • To achieve more uniform image resolution across the field-of-view in CT scans.
  • To mitigate the resolution loss observed in the periphery of reconstructed CT images.

Main Methods:

  • Modeling focal spot angulation effects as local blurrings in transmitted CT intensities.
  • Determining coefficients of discrete convolutions corresponding to these blurrings.
  • Compensating for blurrings in the sinogram domain using a penalized-likelihood restoration model.

Main Results:

  • The proposed method effectively corrects for focal spot angulation-induced blurrings.
  • Uniformity of resolution across the field-of-view in fan-beam CT images is significantly improved.
  • Peripheral image resolution is enhanced, approaching the quality of central regions.

Conclusions:

  • The developed method successfully addresses the challenge of anisotropic resolution in fan-beam CT.
  • Correction for focal spot angulation leads to more diagnostically reliable CT images.
  • This approach offers a practical solution for enhancing overall image quality in CT systems.