Enzyme immunoassay using a reusable extended-gate field-effect-transistor sensor with a
Masao Kamahori1, Yu Ishige, Maki Shimoda
1Central Research Laboratory, Hitachi, Ltd., Kokubunji, Tokyo 185-8601, Japan. masao.kamahori.qj@hitachi.com
Abstract:
A reusable extended-gate field-effect transistor (FET) sensor with an 11-ferrocenyl-1-undecanethiol (11-FUT) modified gold electrode was developed for applying to enzyme immunoassay. It was found that the 11-FUT modified FET sensor detected a thiol compound 50 times or more repeatedly after a treatment with a 5% hydrogen peroxide solution. The gate-voltage shift of the FET sensor showed a fairly good linearity (R(2) = 0.998) within a range from 10(-2) to 10(-6) M on the concentration of 6-hydroxyl-1-hexanethiol, which is a thiol compound, at a Nernstian response of 58.5 mV/decade. The FET-based immunoassay was constructed by combining the 11-FUT modified-FET sensor with the enzyme-linked immunosorbent assay (ELISA), in which the enzyme chemistry of acetylcholinesterase (AChE) was used to generate a thiol compound. The 11-FUT modified FET sensor with an AC voltage at 1 MHz superimposed onto the reference electrode detected the AChE-catalyzed product corresponding to a serum concentration of interleukin 1beta from 10 to 5000 pg/mL. In addition, all measurements were successfully performed by using the same FET-sensor chip after a treatment with a 5% hydrogen peroxide solution.
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Immunogold Electron Microscopy