Related Experiment Video
Updated: Jul 1, 2026

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Accelerating risk of Fidelis lead fracture
David Farwell1, Martin S Green, Robert Lemery
1University of Ottawa Heart Institute, Ottawa, Ontario, Canada.
Background:
Sprint Fidelis (Medtronic Inc., Minneapolis, Minnesota) is a 6.6-F implantable cardioverter-defibrillator lead. In October 2007 the manufacturer suspended distribution of Sprint Fidelis leads secondary to high early fracture rates. Reprogramming of lead impedance alerts was recommended to try to reduce inappropriate shocks.
Objective:
This study sought to assess how the hazard of Fidelis fracture varies with time, whether lead impedance alert thresholds programmed after recall have prevented inappropriate shocks, and predictors of lead fracture.
Methods:
We collected clinical and device interrogation data on all lead fractures and performed a Kaplan-Meier analysis. We performed a case control study to examine univariate and multivariate predictors of lead fracture.
Results:
There have been 17 lead fractures in our cohort of 480 Sprint Fidelis (model 6949) leads, median follow-up 19.8 months. The hazard of fracture increased significantly with time by a power of 2.74 (95% confidence interval: 2.57 to 2.91, P <.0001). Reprogrammed lead impedance thresholds prevented inappropriate shocks in only 1 of 6 patients. Independent predictors of lead fracture were greater left ventricular ejection fraction (P = .0011) and noncephalic venous access (P = .0224).
Conclusion:
The risk of lead fracture increased with time by a power of 2.74. Reprogrammed lead impedance thresholds prevented inappropriate shocks in only 1 of 6 patients. Independent predictors of lead fracture were greater left ventricular ejection fraction and noncephalic access.
More Related Videos
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...
Flail Chest-I
Flail chest is a severe and potentially life-threatening condition characterized by the fracture of three or more adjacent ribs in multiple places. It is most commonly caused by direct impacts and trauma, such as motor vehicle accidents or injuries from a steering wheel impact. It can also occur due to falls in elderly individuals with osteoporosis, or assaults involving sharp objects.
Pathophysiology
The pathophysiology of flail chest is complex, involving fractures of...
Relative Risk