Abstract:
In August 2006, Duke University Perfusion Services had the opportunity to be the first institution in the United States to clinically evaluate the Dideco D100 Neonatal Oxygenator. The device was used on six pediatric patients to facilitate correction or palliation of their cardiac defects, which included two arterial switch operations, two truncus arteriosus repairs, one stage 1 Norwood and one repair of total anomalous pulmonary venous return. The average patient weight was 3.1 kg. The average cardiopulmonary bypass (CPB) time was 135 minutes and the average cross-clamp time was 61 minutes. Arterial and venous blood gasses were drawn and used to calculate oxygen transfer. The average oxygen transfer was 14.8 +/- 10.3 ml/O2/min. The Dideco D100 Oxygenator is the first oxygenation device designed specifically for neonates. The Dideco D100 is a microporous hollow-fiber device. It has a static priming volume of 31 ml and a maximum rated flow of 700 ml/min. The integral hard-shell venous reservoir has a minimum operating level of 10 ml and a reservoir capacity of 500 ml. For this evaluation, the Dideco Kids D100 Neonatal Oxygenator performed adequately on patients weighing up to 5 kg. This device provides an excellent first step towards offering very small children appropriate circuitry without having to sacrifice safety or performance.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...