Related Experiment Video
Updated: Jul 1, 2026

Simultaneous Detection of Different Antibody Classes in a Multiplexed Serological Test
Published on: July 14, 2023
Interference by heterophilic antibodies in immunoassays: wrong increase of myoglobin values
Antonio Bonetti1, Cesare Monica, Chiara Bonaguri
1Dept of Clinical Sciences, University of Parma, Italy. antonio.bonetti@unipr.it
Abstract:
Aim of this work is to illustrate how analytical interference in immunoassay may produce serious errors in clinical laboratory results. The sophisticated quality assurance schemes used in many laboratories do not identify erroneous results arising from aberrant samples. Recently attention has been focused on the incidence and implication of false-positive results arising from the presence of certain substances in a patient's serum that interfere with one or more steps in immunoassays. In this paper, we present the case of a 92 year-old woman whose plasma myoglobin concentrations falsely increased when measured using the Beckman Access assay. We demonstrated that heterophilic antibodies accounted for the falsely increased myoglobin values, and we suggest how to resolve such situations.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...