The reliability of two different cybex isokinetic systems
The Journal of Orthopaedic and Sports Physical Therapy
|January 1, 1990
Summary
This study assessed the reliability of muscular strength measurements using two Cybex II isokinetic dynamometry systems and two data reduction methods. Results show peak torque values from different systems are not always comparable, impacting clinical use.
Area of Science:
- Biomechanics
- Exercise Physiology
- Rehabilitation Science
Background:
- Accurate muscle strength assessment is crucial for various professional fields.
- Isokinetic dynamometry is a common tool for measuring muscular strength.
- Understanding the reliability of different systems and data reduction methods is essential for consistent clinical application.
Purpose of the Study:
- To determine the intertest reliability of muscular strength measurements using two Cybex II isokinetic dynamometry systems.
- To assess the reliability of peak torque values (PTV) obtained from two data reduction methods: Cybex Data Reduction Computers (CDRC) and Cybex Strip Chart Recorders (SCR).
Main Methods:
- Fifty subjects performed knee extension/flexion cycles at two speeds (60 and 240 degrees/sec) over two test days.
- Correlation coefficients and ANOVA were used to analyze the reliability between two Cybex systems and between CDRC and SCR data reduction methods.
- Peak torque values were collected and compared across different testing conditions.
Main Results:
- Strong correlations were found between CDRC and SCR peak torque values within systems (r = 0.987, 0.985) and across systems (r = 0.983, 0.982).
- ANOVA revealed significantly higher CDRC peak torque values from the Cybex II system compared to the Cybex II+ system (p < 0.05).
- No significant differences were found between systems for SCR peak torque values (p > 0.05).
Conclusions:
- Clinicians should exercise caution as Cybex Data Reduction Computer (CDRC) peak torque values may not be comparable across different Cybex II systems.
- Peak torque values obtained using Cybex Strip Chart Recorders (SCR) appear comparable between different Cybex II systems.
- Both CDRC and SCR data reduction methods show comparable results within the same Cybex system.