Validation of the Microlife BP W200-1 wrist device for blood pressure measurement
Paolo Palatini1, Francesca Dorigatti, Elisa Bonso
1Department of Clinical and Experimental Medicine, University of Padova, Padua, Italy. palatini@unipd.it
Objective:
The objective of the study was to determine the accuracy of the BP W200-1 device for blood pressure measurement at the wrist developed by the Microlife Company. The BP W200-1 model features a 'Microlife Average Mode', which uses a 'weighted' average of a minimum of three consecutive blood pressure (BP) readings taken 15 s apart.
Method:
Device evaluation was carried out using the International Protocol of the European Society of Hypertension. Monitor performance was assessed in relation to patients' sex, age, wrist circumference, and systolic (SBP) and diastolic blood pressures (DBP).
Results:
The BP W200-1 comfortably passed all three phases of the protocol for SBP and DBP. The device performed well in all ranges of SBP and DBP. Mean blood pressure differences for the BP W200-1 monitor were -1.2+/-4.5 mmHg for SBP and -0.2+/-3.1 mmHg for DBP. In multivariable analyses, the SBP and DBP discrepancies between device and observer were unrelated to age, sex, wrist circumference, and blood pressure level.
Conclusion:
These data show that the Microlife BP W200-1 wrist monitor satisfies the recommended European Society of Hypertension accuracy levels for SBP and DBP and therefore can be recommended for clinical use in the adult population. Its performance is uniform across subgroups of patients with different clinical characteristics.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure