Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
The design, development, and evaluation of a prototypic, prosthetic venous valve
Matt T Oberdier1, Stanley E Rittgers
1Department of Biomedical Engineering, The University of Akron, Akron, OH 44325, USA.oberdiermt@upmc.edu
Background:
Chronic venous insufficiency is a serious disease for which there is no clearly successful surgical treatment. Availability of a proven prosthetic vein valve could provide such an option by reducing venous reflux while permitting normal antegrade flow.
Methods:
A new prosthetic vein valve design has been developed which mimics the function of a natural valve by ensuring complete closure of the leaflets with minimal obstruction for antegrade flow. A 2:1 mock-up of the device was tested to evaluate its ability to prevent regurgitation and several key modifications were made. A subsequently re-designed 1:1 prototype was then built in 4 slightly different size configurations and then each tested under physiologic conditions of pulsatile flow in both supine and standing positions.
Results:
Each of the configurations showed acceptable amounts of antegrade resistance and effective orifice area and showed low values of regurgitation and % reflux with two of the prototype configurations (flange lengths of 2.5 mm and 3.75 mm) having corresponding values of <2.5 mmHg-min/L, >97%, 11 mL, and 36%, respectively. These values are particularly striking when compared to the corresponding regurgitation and % reflux values of 60 mL and 205%, respectively, when no device is present.
Conclusion:
The results of this study show that this prototype vein valve design is capable of providing significant relief of reflux under realistic conditions without inducing any increase in antegrade flow resistance and warrants further testing with in vivo models.
More Related Videos
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
06:56An Ex Vivo Porcine Model for Hydrodynamic Testing of Experimental Aortic Valve Procedures and Novel Medical Devices
Published on: August 25, 2023
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis III: Interprofessional Care
Varicose Veins II: Diagnostic Studies and Interprofessional Care
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system as it...