Reproducibility of the ambulatory arterial stiffness index in hypertensive patients
Dirk G Dechering1, Marijke S van der Steen, Ahmet Adiyaman
1Department of General Internal Medicine, University Medical Centre St Radboud, Radboud University, Nijmegen, The Netherlands.
Background:
We studied the repeatability of the ambulatory arterial stiffness index (AASI), which can be computed from 24-h blood pressure (BP) recordings as unity minus the regression slope of diastolic on systolic BP.
Methods:
One hundred and fifty-two hypertensive outpatients recruited in Nijmegen (mean age = 46.2 years; 76.3% with systolic and diastolic hypertension) and 145 patients enrolled in the Systolic Hypertension in Europe (Syst-Eur) trial (71.0 years) underwent 24-h BP monitoring at a median interval of 8 and 31 days, respectively. We used the repeatability coefficient, which is twice the SD of the within-participant differences between repeat recordings, and expressed it as a percentage of four times the SD of the mean of the paired measurements.
Results:
Mean AASI (crude or derived by time-weighted or robust regression) and 24-h pulse pressure (PP) were similar on repeat recordings in both cohorts. In Nijmegen patients, repeatability coefficients of AASI and PP were approximately 50%. In Syst-Eur trial patients, repeatability coefficient was approximately 60% for AASI and approximately 40% for PP. For comparison, repeatability coefficients for 24-h systolic and diastolic BP were approximately 30%. Differences in AASI between paired recordings were correlated with differences in the goodness of fit (r2) of the AASI regression line as well as with differences in the night-to-day BP ratio. However, in sensitivity analyses stratified for type of hypertension, r2, or dipping status, repeatability coefficients for AASI did not widely depart from 50 to 60% range.
Conclusion:
Estimates of mean AASI were not different between repeat recordings, and repeatability coefficients were within the 50-60% range.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation