Related Experiment Videos
Evidence for internal filtration in the Genius system, performing slow low efficient daily dialysis in the intensive
Sunny Eloot1, Wim Van Biesen, Annemieke Dhondt
1Renal Division, Department of Internal Medicine, Ghent University Hospital, Ghent, Belgium. sunny.eloot@ugent.be
Background:
Since most intensive care units (ICUs) have no access to high-quality water, the Genius system, using ultrapure water and high-flux dialyzers, might be used during slow low efficient daily dialysis (SLEDD). To get an idea about convective removal, internal filtration (IF) was calculated and compared to other modalities.
Methods:
Validated with in vivo measurements, IF was calculated for 8 h Genius SLEDD and 4 h Genius standard dialysis with high-flux dialyzers, and for 8 h SLEDD and 4 h both with standard dialysis machines and low-flux as well as high-flux dialyzers.
Results:
In Genius SLEDD (FX80), IF was 2.36 l (hematocrit [Hct] 30%) and 3.37 l (Hct 40%). About the same amount of IF was found for the different dialysis modalities, except when using low-flux dialyzers, where no IF occurred.
Conclusion:
The Genius system allows the implementation of high-flux dialysis with ultrapure dialysate, and can be run in a SLEDD mode in the ICU.
Related Concept Videos
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Acute Kidney Injury V: Interprofessional Care
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Hemodialysis I: Introduction
Continuous Renal Replacement Therapy