The effects of flexible bronchoscopy on mechanical ventilation in a pediatric lung model
Danny Hsia1, Robert M DiBlasi2, Peter Richardson1
1Department of Pediatrics, University of Washington School of Medicine, Seattle, WA.
Background:
Flexible bronchoscopy performed through endotracheal tubes (ETTs) in children receiving mechanical ventilation can significantly impact ventilation, but the magnitude of this impact has not been established. We used a lung model to simulate mechanical ventilation in a range of child sizes in order to determine how the insertion of pediatric flexible bronchoscopes into ETTs alters ventilatory parameters, especially tidal volume (Vt) and peak inspiratory pressure (PIP), in both healthy and diseased lungs.
Methods:
We simulated five child sizes based on weight, and evaluated 22 bronchoscope/ETT combinations, first in pressure control (PC) ventilation mode and then in volume control (VC) ventilation mode. The combinations ranged from the 2.2-mm (bronchoscope outer diameter)/3.0-mm (ETT inner diameter) to 5.0-mm bronchoscope/8.0-mm ETT. The primary outcome measures were decrease in Vt after bronchoscope insertion during PC ventilation and increase in PIP during VC ventilation.
Results:
In the PC ventilator mode, Vt decreased by > 50% with nine of the combinations, while during VC ventilation, PIP increased by >or= 20 cm H(2)O with seven combinations. The 2.2-mm bronchoscope/3.0-mm ETT, 2.8-mm bronchoscope/5.0-mm ETT, and 3.6-mm bronchoscope/5.0-mm ETT combinations severely impaired ventilation, while the 3.6-mm bronchoscope/4.5-mm ETT, 5.0-mm bronchoscope/6.5-mm ETT, and 5.0-mm bronchoscope/7.0-mm ETT combinations were incompatible with adequate ventilation.
Conclusions:
The insertion of bronchoscopes into ETTs can lead to clinically relevant decreases in Vt when in the PC ventilator mode and large increases in PIP during VC ventilation. The minimum bronchoscope/ETT diameter difference required to maintain adequate ventilation increases with child size.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation I: Indication and Settings