Related Experiment Video
Updated: Jun 30, 2026

A High Throughput, Multiplexed and Targeted Proteomic CSF Assay to Quantify Neurodegenerative Biomarkers and Apolipoprotein E Isoforms Status
Published on: October 20, 2016
Development and validation of a sensitive LC-MS/MS method with electrospray ionization for quantitation of
D Vijaya Bharathi1, Kishore Kumar Hotha, P V Vidya Sagar
1Bioanalytical Department, Integrated Product Development, Dr Reddy's Laboratories Ltd, Bachupalli, Hyderabad-500 072, India.
Abstract:
A highly sensitive and specific LC-MS/MS method has been developed and validated for the estimation of pramipexole (PPX) with 500 microL human plasma using memantine as an internal standard (IS). The API-4000 was operated under multiple-reaction monitoring mode (MRM) using the electrospray ionization technique. Solid-phase extraction was used to extract PPX and IS from human plasma. The resolution of peaks was achieved with 0.01 m ammonium acetate buffer (pH 4.4):acetonitrile (30:70, v/v) on a Discovery CN column. The total chromatographic run time was 3.0 min and the elution of PPX and IS occurred at approximately 2.32 and 2.52, respectively. The MS/MS ion transitions monitored were 212.10 --> 153.10 for PPX and 180.20 --> 107.30 for IS. The method was proved to be accurate and precise at linearity range of 20-3540 pg/mL with a correlation coefficient (r) of > or =0.999. The intra- and inter-day precision and accuracy values found to be within the assay variability limits as per the FDA guidelines. The developed assay method was applied to a pharmacokinetic study in human volunteers following oral administration of 0.25 mg PPX tablet.