Related Experiment Video
Updated: Jun 30, 2026

A Polyaniline-based Sensor of Nucleic Acids
Published on: November 1, 2016
Polyaniline-carbon nanotube composite film for cholesterol biosensor
Chetna Dhand1, Sunil K Arya, Monika Datta
1Biomolecular Electronics and Conducting Polymer Research Group, National Physical Laboratory, Dr. K. S. Krishnan Marg, New Delhi 110012, India.
Abstract:
Nanocomposite film composed of polyaniline (PANI) and multiwalled carbon nanotubes (MWCNT), prepared electrophoretically onto indium tin oxide (ITO)-coated glass plate, was used for covalent immobilization of cholesterol oxidase (ChOx) via N-ethyl-N'-(3-dimethylaminopropyl) carbodiimide (EDC) and N-hydroxysuccinimide (NHS) chemistry. Results of linear sweep voltammetric measurements reveal that ChOx/PANI-MWCNT/ITO bioelectrode can detect cholesterol in the range of 1.29 to 12.93 mM with high sensitivity of 6800 nA mM(-1) and a fast response time of 10 s. Photometric studies for ChOx/PANI-MWCNT/ITO bioelectrode indicate that it is thermally stable up to 45 degrees C and has a shelf life of approximately 12 weeks when stored at 4 degrees C. The results of these studies have implications for the application of this interesting matrix (PANI-MWCNT) toward the development of other biosensors.
More Related Videos
11:18Manufacturing of a Nafion-coated, Reduced Graphene Oxide/Polyaniline Chemiresistive Sensor to Monitor pH in Real-time During Microbial Fermentation
Published on: January 7, 2019
08:06The Use of a β-lactamase-based Conductimetric Biosensor Assay to Detect Biomolecular Interactions
Published on: February 1, 2018