Objectives:
Arterial oxygen saturation (SaO(2)) monitoring using pulse oximeter (SpO(2)) is mandatory in the intensive care unit. The aim was to assess bias and precision of new (SpO(2)ng) and old (SpO(2)og) pulse oximeter technologies in the postoperative period following pediatric cardiac surgery in cyanotic children.
Study Design:
Prospective, monocentric.
Patients And Methods:
Ten patients (7 days to 53 months old) were studied in the postoperative period following palliative cardiac surgery. SaO(2), SpO(2)og, and SpO(2)ng were obtained every 4 hours. SaO(2) of arterial blood sample was obtained from an intra-arterial catheter located in the radial artery, on the same side as the oximeters. Bias and precision were assessed using Bland-Altman analysis.
Results:
We obtained 136 SaO(2) determinations. Mean SaO(2) was 76+/-15%. SpO(2)og was significantly different from SaO(2), while SpO(2)ng was not different from SaO(2). In 21 (15%) cases, SpO(2)og was not available whereas SpO(2)ng was available in 136 (100%) cases. In the remaining 115 cases, SpO(2)ng's precision was significantly better than SpO(2)og's precision.
Discussion:
SpO(2)ng is more accurate and more reliable than SpO(2)og for SaO(2) monitoring in the postoperative period following pediatric cardiac surgery in cyanotic children.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
07:51Standardized Hemorrhagic Shock Induction Guided by Cerebral Oximetry and Extended Hemodynamic Monitoring in Pigs
Published on: May 21, 2019
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...