Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
Validation of volume-pressure recording tail-cuff blood pressure measurements
Minjie Feng1, Steven Whitesall, Yunyu Zhang
1Novartis Institutes for BioMedical Research & Novartis Pharmaceutical Corporation, East Hanover, NJ, USA.
Background:
The American Heart Association has recommended tail-cuff blood pressure measurement for high-throughput experimental designs, including mutagenesis screens and genetic crosses. However, some tail-cuff methods show good agreement with radiotelemetry and others do not, indicating that each tail-cuff method requires independent validation.
Methods:
We validated the volume-pressure recording (VPR) tail-cuff method by comparison to simultaneous radiotelemetry measurements.
Results:
Bland-Altman analysis of 560 cycles from 26 independent measurement sessions showed good agreement between VPR and radiotelemetry measurements, with tail-cuff measurements being 0.25 mm Hg lower than telemetry measurements on average. However, the VPR method was less accurate, compared to radiotelemetry, at extreme high and low (i.e., <110 or >180 mm Hg) systolic blood pressures (SBPs).
Conclusions:
We conclude that the VPR tail-cuff method provides accurate blood pressure measurements over the physiological range of blood pressure in mice.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Assessing Blood pressure in the Leg
Preparation: