Three-dimensional analysis of computed tomography-based navigation system for total knee arthroplasty: the accuracy
Hideki Mizu-Uchi1, Shuichi Matsuda, Hiromasa Miura
1Department of Orthopaedic Surgery, Graduate School of Medical Sciences, Kyushu University, Fukuoka City, Japan.
Abstract:
We evaluated the postoperative alignment of 37 primary total knee arthroplasties performed using a computed tomography-based navigation system (Vector Vision Knee 1.5; Brain Lab, Germany) with a new 3-dimensional analysis. The mean coronal femoral angle was 89.0 degrees +/- 1.4 degrees (85.5 degrees -92.8 degrees ), and the coronal tibial component was 89.2 degrees +/- 1.0 degrees (87.4 degrees -91.6 degrees ). The hip-knee-ankle angle was observed to be 178.2 degrees +/- 1.5 degrees (173.9 degrees -181.8 degrees ). The external rotational alignment of the femoral component relative to the surgical epicondylar axis was -0.5 degrees +/- 1.7 degrees (-3.2 degrees to 3.4 degrees ). The results demonstrated that a computed tomography-based navigation system provided a reasonably satisfactory component alignment. The discrepancy between the 2-dimensional and 3-dimensional evaluations was 1.0 degrees +/- 0.9 degrees (0.1 degrees -3.4 degrees ). Three-dimensional analysis is necessary to evaluate the accuracy of the navigation system.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography