The relationship between increased hip range of motion, wear, and locking mechanism failure in the Harris-Galante

Hiroshi Imai1, Naohiko Mashima, Toshiaki Takahashi

  • 1Department of Bone and Joint Surgery, Ehime University Graduate School of Medicine, Ehime, Japan.

We performed both clinical and radiographic evaluations of 178 patients (190 hips) who had undergone cementless total hip arthroplasties using Harris-Galante I/II porous cups after an average 12-year follow-up period (range, 8-18 years). We revised 15 Harris-Galante I/II porous cups (7.8%), and the locking mechanism was broken in 10 revised cups (67%). There was a significant association between locking mechanism failure and linear polyethylene wear. We observed a significant positive correlation between linear polyethylene wear and increased ranges of motion such as flexion, adduction, and external rotation at the last follow-up visit after the primary operation. Increased ranges of motion seen in Asians induced higher linear polyethylene wear and locking mechanism failure due to impingement of the neck and cup.