Related Experiment Video
Updated: Jun 29, 2026

NiO Nanoflowers for Non-Enzymatic Amperometric Detection of Glucose
Published on: December 30, 2025
Fabrication of a glucose sensor based on a novel nanocomposite electrode
Afsaneh Safavi1, Norouz Maleki, Elaheh Farjami
1Department of Chemistry, College of Sciences, Shiraz University, Shiraz, Iran. safavi@chem.susc.ac.ir
Abstract:
The direct electrocatalytic oxidation of glucose in alkaline medium at nanoscale nickel hydroxide modified carbon ionic liquid electrode (CILE) has been investigated. Enzyme free electro-oxidation of glucose have greatly been enhanced at nanoscale Ni(OH)(2) as a result of electrocatalytic effect of Ni(+2)/Ni(+3) redox couple. The sensitivity to glucose was evaluated as 202 microA mM(-1)cm(-2). From 50 microM to 23 mM of glucose can be selectively measured using platelet-like Ni(OH)(2) nanoscale modified CILE with a detection limit of 6 microM (S/N=3). The nanoscale nickel hydroxide modified electrode is relatively insensitive to electroactive interfering species such as ascorbic acid (AA), and uric acid (UA) which are commonly found in blood samples. Long-term stability, high sensitivity and selectivity as well as good reproducibility and high resistivity towards electrode fouling resulted in an ideal inexpensive amperometric glucose biosensor applicable for complex matrices.