Readily synthesized calibration compounds for quadrupole and magnetic instruments for use over the mass range to 2000
F F Hsu1, A N Tyler, W R Sherman
1Department of Medicine, Washington University, School of Medicine, St Louis, Missouri 63110.
Abstract:
Volatile mass calibration standards have been prepared by esterifying beta-cellobiose (4-beta-D-glucopyranosyl-D-glucose) with a mixture of heptafluorobutyric (HFB) anhydride and pentafluoropropionic (PFP) anhydride. The mixed esters produce spectra that are useful for mass spectrometer calibration with positive or negative ion methane chemical ionization and electron impact over the mass range 300-2000. The spectra contain prominent ions spaced at m/z 20 and m/z 50 intervals. Using the mixed ester with direct insertion probe introduction gives intense spectra that persist for tens of minutes. All signals above m/z 194 derived from these substances disappear rapidly upon withdrawal of the probe. The composition and exact masses are given for the positive and negative ion spectra of a mixed HFB/PFP ester of beta-cellobiose. Two other calibrants are described: one made from beta-cellobiose using a mixture of HFB, PFP and trifluoroacetic anhydrides, and another the HFB/PFP mixed ester of perseitol. These are examples of the flexibility of this approach with respect to mass range and ion composition.
More Related Videos
06:56Characterization of Synthetic Polymers via Matrix Assisted Laser Desorption Ionization Time of Flight (MALDI-TOF) Mass Spectrometry
Published on: June 10, 2018
09:38Single-throughput Complementary High-resolution Analytical Techniques for Characterizing Complex Natural Organic Matter Mixtures
Published on: January 7, 2019
Related Concept Videos
High-Resolution Mass Spectrometry (HRMS)
MALDI-TOF Mass Spectrometry
Mass Analyzers: Overview
Mass Analyzers: Common Types
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...