Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
Arterial pulse cardiac output agreement with thermodilution in patients in hyperdynamic conditions
Giorgio Della Rocca1, Maria Gabriella Costa, Paolo Chiarandini
1Department of Anesthesia and Intensive Care Medicine, University of Udine, Udine, Italy.
Objective:
This study aimed to compare continuous cardiac output (CCO) obtained using the arterial pulse wave (APCO) measurement with a simultaneous measurement of the intermittent cardiac output (ICO) and CCO obtained with a pulmonary artery catheter (PAC) in liver transplant patients.
Design:
A prospective, single-center evaluation.
Setting:
A university hospital intensive care unit.
Patients:
Eighteen patients after liver transplantation.
Interventions:
Pulmonary artery catheters were placed in all patients, and ICO and CCO were determined using thermodilution. APCO measurements were made with the Vigileo System (Edwards Lifesciences, Irvine, CA).
Measurements And Main Results:
The authors obtained 126 data pairs of ICO and APCO and 864 pairs of CCO and APCO. ICO data were collected after intensive care unit admission and every 8 hours until the 48th postoperative hour. CCO and APCO data were collected every hour from admission until the 48th postoperative hour. Bias and precision were 0.95 +/- 1.41 L/min for ICO versus APCO and 1.29 +/- 1.28 L/min for CCO and APCO. Bias and precision for cardiac output (CO) data pairs less than 8 L/min were 0.32 +/- 1.14 L/min between ICO and APCO and 0.71 +/- 0.98 L/min between CCO and APCO. For CO data pairs higher than 8 L/min, bias and precision were 1.79 +/- 1.54 L/min between ICO and APCO and 2.25 +/- 1.14 L/min between CCO and APCO.
Conclusions:
APCO enables the assessment of CO with clinically acceptable bias and precision. At higher CO levels, APCO underestimates PAC measurements and it is not as reliable as thermodilution in hyperdynamic liver transplant patients.
Related Concept Videos
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Pulse
The pulse serves as a clinical indicator...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...