Evaluation of a random access analyser: BM/Hitachi 911
T Kanluan1, S Tangvarasittichai, O Tangvarasittichai
1Department of Clinical Chemistry Faculty of Medical Technology Mahidol University Siriraj Hospital Bangkok 10700 Thailand.
Abstract:
The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtainedfrom BM/Hitachi 911 and 717 were between the extreme values of 0.94 and 1.05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation.