An evaluation of the Kemtek 1000 sample processor
Related Experiment Video
Updated: Jun 28, 2026

Fruit Volatile Analysis Using an Electronic Nose
Published on: March 30, 2012
1Department of Chemical Pathology St Thomas' Hospital London SE1 7EH UK.
Abstract:
The Kemtek 1000 Sample Processor has been evaluated for precision, accuracy, speed and reliability. Precision was better than 1.0% at all volumes tested and accuracy within +/-5%. A l00-tube assay could be set up within 15 min when patient specimens plus two reagents were sampled using a two probe system. Carry-over could be reduced to <0.01% by using a sufficient number of wash steps, the latter being related to the assay requirements. Evidence was found for adsorption of protein to the probe tubing but inaccuracies due to this could be reduced by introducing wash steps between samples. Problems over 12 months have been minor and quickly resolved. The authors were pleased with the way the processor performed and their staffhave confidence in leaving it to set up their assays.
Related Concept Videos
Sample Preparation for Analysis: Overview
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Sample Handling
Samples should be transported carefully from collection points to the laboratory. They should be properly sealed and clearly labeled to prevent cross-contamination. To preserve the sample integrity, optimal temperature conditions during transport are essential. This could involve using...