Determination of inorganic phosphorus in serum: Evaluation of three methods applied to the Technicon RA-1000 analyzer
L Rubino1, V Catapano, G Guerra
1Clinical Chemistry Laboratory 'Borselli' Hospital (USL 31) Bondeno Ferrara 44012 Italy.
Abstract:
We have evaluated three analytical methods for determining inorganic phosphorus in serum applied to the Technicon RA-I000 analyzer: a fully enzymatic colorimetric method based on the specific system purine nucleoside phosphorylase/xanthine oxidase coupled to an indicator colorimetric reaction similar to the Trinder reaction; a chemical method involving the direct UV measurement of the phosphomolybdate complex; and a chemical method with reduction of the phosphomolybdate complex to molybdenum blue. Experiments were performed to assess within-run and between-day precision, linearity, interference and correlation. The best performance characteristics were shown by the enzymatic colorimetric method and the phosphomolybdate UV method.