Evaluation of the Dacos 3.0 analyser
1Servei d'Anàlisis Clíniques Centre d'Assistència Primàlria 'Dr. Robert' Plaça de la Medicina Badalona Barcelona s/n 08915 Spain.
The Journal of Automatic Chemistry
|January 1, 1990
Summary
The Coulter Dacos 3.0 analyser demonstrates acceptable photometric performance and satisfactory precision in routine clinical laboratory tests. Minor reagent carry-over and interferences were noted, but overall accuracy was good for most analytes.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Laboratory Instrumentation
Background:
- The evaluation of automated clinical laboratory analyzers is crucial for ensuring reliable diagnostic testing.
- Standardized guidelines from organizations like the European Committee for Clinical Laboratory Standards (ECCLS) are essential for performance assessment.
Purpose of the Study:
- To comprehensively evaluate the analytical performance and practicability of the selective multitest Coulter Dacos 3.0 analyser.
- To assess the analyzer's accuracy, imprecision, linearity, drift, and susceptibility to interferences.
Main Methods:
- Photometric evaluation at 340 nm and 420 nm, including inaccuracy and imprecision assessments.
- Routine working evaluation of seven analytical methods (glucose, creatinine, iron, total protein, AST, ALP, calcium) for within-run and between-run imprecision.
- Study of sample and reagent carry-over, and assessment of interferences from turbidity and hemolysis.
Main Results:
- Acceptable inaccuracy and low imprecision (0.06-0.28%) in photometric measurements.
- Satisfactory imprecision for pipette delivery, temperature control, and washing systems.
- Within-run imprecision ranged from 0.6% to 7.6% (CV), and between-run imprecision from 1.2% to 15.1% (CV) across various analytes and concentrations.
- No sample carry-over detected; minor reagent carry-over observed for total protein.
- Weak interference from turbidity for LDH and urate; negative interference from hemolysis for gamma-GT, amylase, bilirubin, and ALP.
Conclusions:
- The Coulter Dacos 3.0 analyser exhibits good overall analytical performance for routine clinical chemistry assays.
- Minor interferences and reagent carry-over require consideration but do not significantly compromise the analyzer's utility.
- The analyzer is deemed suitable for routine laboratory use based on its performance characteristics.
Related Concept Videos
Mass Analyzers: Overview
The mass analyzer is a crucial component of the mass spectrometer. In the ionization chamber, the vaporized sample is bombarded with a high-energy electron beam to generate a radical cation and further fragment into neutral molecules, radicals, and cations. A series of negatively charged accelerator plates accelerate the cations into the mass analyzer. The mass analyzer separates ions according to their mass-to-charge (m/z) ratios and then directs them to the detector. The common types of mass...
Development of Analytical Methods
An analytical methodology can be divided into four sequential steps: technique, method, procedure, and protocol. A technique is a scientific principle that rationalizes a specific phenomenon through chemical measurements. Adapting a technique for analyzing a sample of interest is termed a method. The procedure outlines the directions for performing the analysis via an analytical method. The protocol is the detailed guidelines on the procedure, which should be strictly followed to obtain the...
Data Validation
Method validation is a crucial process in analytical chemistry designed to confirm that a given method consistently produces reliable and high-quality results. This process is essential when a method is applied to different sample matrices or when procedural modifications are made, ensuring that the results meet acceptable standards across various applications.
Key parameters for method validation include:
Key parameters for method validation include:
Mass Analyzers: Common Types
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
Qualitative Analysis
For solutions containing mixtures of different cations, the identity of each cation can be determined by qualitative analysis. This technique involves a series of selective precipitations with different chemical reagents, each reaction producing a characteristic precipitate for a specific group of cations. Metal ions within a group are further separated by varying the pH, heating the mixture to redissolve a precipitate, or adding other reagents to form complex ions.
For instance, group IV...
For instance, group IV...
Qualitative Analysis
Qualitative analysis is the process of identifying elements, ions, or compounds in an unknown sample. It is the first and most fundamental type of analysis based on the hierarchy of analytical goals. This hierarchy is significant as it provides a structured approach to scientific research, with qualitative analysis serving as the initial step, providing essential information before moving on to quantitative or other forms of analysis.
There are two main approaches to qualitative analysis:...
There are two main approaches to qualitative analysis:...