Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Point-of-care bedside gas analyzer: limited use of venous pCO2 in emergency patients
Irwani Ibrahim1, Shirley B S Ooi, Chan Yiong Huak
1Department of Emergency Medicine, National University Hospital, Singapore.
Background:
Because arterial punctures are more painful, venous blood gas analysis has been proposed as an alternative.
Objectives:
To determine if venous pCO(2) can replace arterial pCO(2) in emergency patients using a bedside blood gas analyzer.
Methods:
This is a cross-sectional study. We recruited patients who were deemed by the attending Emergency Physicians to require arterial blood gas analysis to determine their ventilation or acid-base status. A venous and an arterial blood gas sample were drawn from the patient, temporally as close to each other as possible. Both samples were then analyzed using the same bedside blood gas analyzer immediately after collection.
Results:
There were 122 paired samples obtained. The strength of the association between arterial and venous pCO(2) is r = 0.838 (p = 0.001). The Bland-Altman bias plot methods for agreement show a mean difference of 3.3 mm Hg with two standard deviation limits of agreement being -17.4 to 23.9; 93.4% of the pCO(2) values fell within two standard deviation limits. Venous pCO(2) below 30 mm Hg had a 100% (95% confidence interval [CI] 90.5-100) sensitivity and 100% (95% CI 80.7-100) Negative predictive value to rule out hypercarbia, defined as arterial pCO(2) > 45 mm Hg.
Conclusions:
There is significant variability in the arteriovenous difference of pCO(2) readings, hence, arterial punctures are still needed to specifically determine the arterial pCO(2).
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and HCO3− values are examined to...
Chemical Factors Affecting Respiration Centers
CO2 has a potent influence on respiration and is strictly regulated. Under...
Oxygen Transport in the Blood
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...