Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Evaluation of flow scintillation analysis for the determination of Sr-90 in bioassay samples
Alexander A Plionis1, Edward R Gonzales, Sheldon Landsberger
1Chemistry Division, Los Alamos National Laboratory, Mail Stop K484, Los Alamos, NM 87545, USA.
Abstract:
An automated procedure for the determination of (90)Sr was adapted from existing methods of flow scintillation analysis (FSA) for use on aqueous samples with low levels of activity (<1000 dpm per sample). This technique employs high-performance extraction chromatography (HPEC) and an on-line liquid scintillation counter to provide automated separation and subsequent detection of (90)Sr. The total analysis time is 30 min per sample. Dilute urine samples, spiked with (90)Sr, were also processed by this method to test the application of this technique for bioassay monitoring.