Comparison of 3 optical navigation systems for computer-aided maxillofacial surgery
E Bradley Strong1, Amir Rafii, Bettina Holhweg-Majert
1Department of Otolaryngology, University of California, Davis, School of Medicine, 2521 Stockton Blvd, Ste 7200, Sacramento, CA 95817, USA. edward.strong@ucdmc.ucdavis.edu
The StealthStation navigation system demonstrated the highest accuracy in maxillofacial reconstruction among three computer-aided surgery systems evaluated. However, observed accuracy differences were clinically negligible for surgical applications.
Area of Science:
- Maxillofacial Surgery
- Computer-Aided Surgery
- Surgical Navigation Systems
Background:
- Computer-aided surgery (CAS) systems are increasingly utilized in complex reconstructive procedures.
- Accurate navigation is crucial for optimizing outcomes in maxillofacial reconstruction.
Purpose of the Study:
- To compare the accuracy of three distinct CAS systems for maxillofacial reconstruction.
- To identify the most accurate system among StealthStation, VectorVision, and Voxim.
Main Methods:
- Evaluation of StealthStation, VectorVision, and Voxim CAS systems.
- Testing conducted using four fresh cadaveric heads in a dedicated laboratory setting.
- Mean target registration error was the primary metric for accuracy assessment.
Main Results:
- StealthStation exhibited the highest accuracy with a mean target registration error of 1.00 mm.
- VectorVision (1.13 mm) and Voxim (1.34 mm) showed progressively lower accuracy.
- All observed differences in accuracy were statistically significant (P < .05).
Conclusions:
- Significant accuracy variations exist among the evaluated CAS navigation systems.
- StealthStation was found to be the most accurate system in this study.
- Despite statistical significance, the small magnitude of accuracy differences suggests negligible clinical impact for maxillofacial reconstruction.