A non-contact vital sign monitoring system for ambulances using dual-frequency microwave radars
Satoshi Suzuki1, Takemi Matsui, Hiroshi Kawahara
1Department of Management Systems Engineering, Tokyo Metropolitan University, 6-6 Asahigaoka, Hino, Tokyo 191-0065, Japan.
Abstract:
We developed a novel non-contact monitoring system to measure the vital signs of casualties inside a moving ambulance. This system was designed to prevent exposure of patients to infectious organisms under biochemical hazard conditions. The system consists of two microwave radars: a 10-GHz respiratory-monitoring radar is positioned 20 cm away from the surface of the isolator. The 24-GHz cardiac-monitoring radar is positioned below the stretcher underneath the isolator. The subject (22.13 +/- 0.99 years) was placed inside the isolator on a stretcher in the simulated ambulance. While the ambulance was in motion at a speed of approximately 10 km/h, the heart rates determined by the cardiac-monitoring radar correlated significantly with those measured by ECG (r = 0.69, p < 0.01), and the respiratory rates derived from the respiratory-monitoring radar correlated with those measured by the respiration curves (r = 0.97, p < 0.0001). The proposed system appears promising for future on-ambulance monitoring of the vital sign of casualties exposed to toxins.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Holter Monitor: 24-Hour Monitoring
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Doppler Effect - II