Technical feasibility of transperineal MR-guided prostate interventions in a low-field open MRI unit: canine study

Ferenc Lakosi1, Gergely Antal, Csaba Vandulek

  • 1Health Science Center, University of Kaposvar, Guba Sandor Street 40, Kaposvar H-7400, Hungary.

Insights

This study shows a low-field (0.35 T) transperineal system is feasible for precise Magnetic Resonance Imaging (MRI) guided prostate interventions in canines. The system achieved accurate needle placement, proving promising for future prostate cancer treatments.

Area of Science:

  • Urology
  • Medical Imaging
  • Interventional Radiology

Background:

  • Magnetic Resonance Imaging (MRI) offers excellent visualization of prostate anatomy and lesions.
  • Accurate guidance is crucial for effective prostate interventions like brachytherapy.

Purpose of the Study:

  • To demonstrate the feasibility of a low-field (0.35 T) transperineal system for precise MRI-guided prostate interventions.
  • To evaluate the accuracy and reliability of this system in a canine model.

Main Methods:

  • Utilized T2-weighted FSE images for planning and fast gradient spoiled-echo (FSPGR) sequences for real-time guidance.
  • Employed MR-compatible coaxial needles inserted transperineally for brachytherapy catheter placement.
  • Validated needle placement accuracy through pathology examination.

Main Results:

  • Achieved a mean needle displacement deviation of 2.9 mm (median 2.7 mm), with 97% of errors < 4.0 mm.
  • Needle placement accuracy was modeled by Rayleigh distribution (sigma = 2.3 mm).
  • Demonstrated feasibility within a reasonable time frame (total procedure time approx. 2.5 hours).

Conclusions:

  • The low-field transperineal MRI guidance system is a feasible, accurate, and reliable approach for prostate interventions.
  • This method shows promise for high-quality, time-efficient prostate MR guidance.