Related Experiment Video
Updated: Jun 28, 2026

Guidelines for Elective Pediatric Fiberoptic Intubation
Published on: January 17, 2011
Evaluation of the pediatric Bonfils fiberscope for elective endotracheal intubation
Berthold Bein1, Frank Wortmann, Patrick Meybohm
1Department of Anaesthesiology and Intensive Care Medicine, University Hospital Schleswig-Holstein, Campus Kiel, Germany. bein@anesthesie.uni-kiel.de
Insights
The pediatric Bonfils fiberscope showed a high failure rate and prolonged intubation times in children, indicating significant drawbacks for routine pediatric endotracheal intubation. Further research is needed to improve its efficacy.
Area of Science:
- Anesthesiology
- Pediatric Medicine
- Medical Devices
Background:
- Pediatric difficult airway management presents unique challenges for healthcare providers.
- The study evaluates the performance of new pediatric Bonfils fiberscope versions for endotracheal intubation.
Purpose of the Study:
- To assess the efficacy and safety of the pediatric Bonfils fiberscope for elective endotracheal intubation in children.
- To determine the success rate and time required for intubation using this device.
Main Methods:
- 55 children undergoing elective surgery were enrolled after IRB approval and informed consent.
- Two pediatric Bonfils fiberscope sizes (2 mm and 3.5 mm OD) were used based on endotracheal tube size.
- Children were divided into two groups: 19 with atropine pretreatment and 36 without.
Main Results:
- First-attempt success rates were 69% (without atropine) and 78% (with atropine).
- Overall failure rates after three attempts were 14% (without atropine) and 5% (with atropine).
- Intubation failures in both groups were primarily due to secretions obscuring the optic aperture.
Conclusions:
- The pediatric Bonfils fiberscope demonstrated a high failure rate and extended intubation times in normal pediatric airways.
- These findings suggest significant limitations for its routine use in pediatric endotracheal intubation.
- Secretions interfering with the optic aperture represent a key challenge for this device.
Background:
Difficult airway management in children is a particular challenge for anesthesiologists and pediatricians. This study was designed to evaluate the performance of the recently developed pediatric versions of the Bonfils fiberscope for elective endotracheal intubation during routine surgical procedures.
Methods:
After approval by the institutional review board and written informed consent, 55 children (age 6 +/- 4 years) scheduled for elective minor surgical procedures were enrolled. Nineteen children received atropine before the intubation attempt, while in the remaining 36 children, no antisialogogue was used. For endotracheal tubes up to 3.5 mm internal diameter, a fiberscope with outer diameter (OD) 2 mm, and for larger endotracheal tubes, a fiberscope OD 3.5 mm was used. Time to intubation and failure rate were obtained.
Results:
In the 36 children without and the 19 children with atropine pretreatment, the success rate for tracheal intubation on the first attempt was 69%/78% (25/15 patients). 4/3 patients and 2/0 patients were intubated after two and three attempts, respectively, and in 5/1 patients (14%/5%) intubation failed even after three attempts. Time to intubation was median 58/60 s, 25th-75th percentile 35-100/32-110 s, and range 14-377/18-360 s. In both groups, failed intubations were because of the secretions contaminating the optic aperture.
Conclusions:
High failure rate and increased intubation times suggest that the pediatric Bonfils fiberscope has significant drawbacks when used for intubation of normal pediatric airways.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...