Coulometric trace determination of water by using Karl Fischer reagent and potentiometric end-point detection
1Department of Analytical Chemistry, University of Umeå, 901 87 Umeå, Sweden.
Abstract:
A new approach to the determination of water via the Karl Fischer reaction is described. Iodine is coulometrically generated and the end-point corresponding to a slight excess of iodine, is detected potentiometrically with a non-polarized platinum electrode. Samples of 1-500 mul containing 0.05-200 mug of water were analysed with a standard deviation of 0.015 mug in the range 0.05-20 mug of H(2)O. A specially constructed electrolysis cell was used in combination with an LKB 16300 Coulometric Analyzer and the time for a complete analysis was 1-4 min, depending on sample size. The reagent composition has been optimized in order to enhance the rate of the main reaction and to minimize the extent of side-reactions. Decreasing the temperature reduced the extent of side-reactions. The displacement of end-point potential on dilution was studied and a correction is discussed.
Related Concept Videos
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Potentiometric Titration: Overview
Controlled-Current Coulometry: Overview
Coulometry: Overview
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Controlled-Current Coulometry: Coulometric Titration
The fundamental requirements for coulometric titrations are (1) 100% efficiency in the reagent-generating electrode reaction and (2) a stoichiometric and preferably rapid reaction between the generated...