A statistical approach by serial correlation to assessment of stability of analytical systems
1Department of Analytical Chemistry, University of Cluj-Napoca, Cluj-Napoca, Romania.
Abstract:
The functional instability of infrared spectrometers can result in serious errors of analysis. This work shows that the stability can be tested by means of the serial correlation method, and the various causes of unstable conditions and equipment faults identified. Drift and large oscillations present separately or together can be detected by new statistics developed from the serial correlation method.
Related Concept Videos
Statistical Analysis: Overview
One of the most commonly used statistical quantifiers is the mean, which is the ratio between the sum of the numerical values of all results and the...
Data Validation
Key parameters for method validation include:
Correlation and Regression
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Stability of structures
Development of Analytical Methods