Related Experiment Video
Updated: Jun 28, 2026

An Anaerobic Biosensor Assay for the Detection of Mercury and Cadmium
Published on: December 17, 2018
Effect of container material, storage time and temperature on determinations of cadmium levels in human urine
K S Subramanian1, J C Meranger, J Connor
1Environmental Health Directorate, Health and Welfare Canada, Tunney's Pasture, Ottawa, Ontario, Canada.
Abstract:
The effect of container material (polyethylene, polypropylene, polystyrene, borosilicate glass and flint glass), storage time (0-86 days), and temperature (22 degrees and 4 degrees ) on the cadmium content found by graphite-furnace atomic-absorption spectrometry for spiked and unspiked urine samples has been studied. No loss of cadmium occurred for at least 10 and 28 days when unspiked and spiked samples respectively, were stored in polyethylene containers at 22 degrees . For storage up to 3-4 days at 22 degrees , polypropylene, polystyrene, borosilicate glass and flint glass containers were also found suitable. Storage at 4 degrees was not effective in preventing loss of cadmium.
Related Concept Videos
Urine Studies I: Urinalysis
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...