Related Experiment Video
Updated: Jun 28, 2026

Extending the Lifespan of Soluble Lead Flow Batteries with a Sodium Acetate Additive
Published on: January 7, 2019
Anodic stripping voltammetric determination of trace lead with a nafion/crown-ether film electrode
1Changchun Institute of Applied Chemistry, Academia Sinica, Changchun, Jilin 130021, People's Republic of China.
Abstract:
An electrode for the anodic stripping voltammetric determination of trace lead has been made by coating glassy carbon with a film of Naflon in which a crown ether is incorporated. The sensitivity is increased because of continuous transfer of lead from solution to the electrode surface by complexation with the crown ether and reduction during the deposition period. A detection limit of 5 x 10(-10)M has been obtained by electrodeposition for 3 min. The method is sensitive, simple and relatively rapid, with a relative standard deviation of 8% at the 2 x 10(-9)M level.
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Electrodeposition
Electrodeposition can...
Electrogravimetric Analysis: Overview
To test the completeness of the...
Voltammetric Techniques: Linear-Scan (E vs Time)
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...