Related Experiment Video
Updated: Jun 28, 2026

Preparation of Food Samples Using Homogenization and Microwave-Assisted Wet Acid Digestion for Multi-Element Determination with ICP-MS
Published on: December 22, 2023
Determination of lead in tooth-pastes by electrothermal atomic-absorption spectrophotometry with platform atomization
Z A Khammas1, M H Farhan, M M Barbooti
1Central Organization for Standardization and Quality Control, P.O. Box 13032, Jadiriyah, Baghdad, Iraq.
Abstract:
Electrothermal atomic-absorption spectrometry is used for the determination of lead in tooth-pastes, by means of the graphite platform/matrix-modification technique. The method is easy to apply, has limited interferences and is more precise than conventional flame AAS. The working concentration range is up to 100 mug/1. The method is characterized by a detection limit of 0.08 mug/1. and a precision of +/-1.9 % for the sample solutions. For the tooth-pastes, the detection limit is between 1.1 and 5.4 ng/g and the precision in the range 1.8-10%.
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...