Flow-injection determination of phosphate with a cadmium ion-selective electrode
D E Davey1, D E Mulcahy, G R O'Connell
1School of Chemical Technology, S.A. Institute of Technology, P.O. Box 1, Ingle Farm, South Australia 5098.
Abstract:
A flow-injection method is described, in which phosphate standards are introduced into a reagent stream containing Cd(2+) ,resulting in the formation of Cd(3)(PO(4))(2). The associated reduction in free metal concentration is sensed by a cadmium-selective electrode. With the exception of major interference from iodide and moderate interference from bromide and thiocyanate, the system exhibits excellent response to phosphate and selectivity over several common anions in solutions buffered at pH 8.4. A maximum sampling rate of 160/hr is possible for phosphate standards in the concentration range 10(-1)-10(-1)M with a 10(-4)M Cd(2+) reagent stream at a total flow-rate (carrier and reagent stream combined) of 8.4 ml/min.
Related Concept Videos
Electrodeposition
Electrodeposition can...
Factors Affecting Solubility
Gas Chromatography: Types of Detectors-II
Washing, Drying, and Ignition of Precipitates
Qualitative Analysis
For instance, group IV...
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides and...