Accuracy in the precise coulometric titration of ammonia and ammonium ion with electrogenerated hypobromite
1National Chemical Laboratory for Industry, 1-1, Higashi, Tsukuba-shi, Ibaraki 305, Japan.
Abstract:
A precise and accurate method in which ammonia and ammonium ions are coulometrically titrated with electrogenerated hypobromite was studied. Through the measurement of the current efficiency for generating hypobromite and the dependencies of titration results on electrolysis current and delay time between sampling and starting electrolysis, precision and accuracy of 0.01 and 0.02%, respectively, were obtained. The accuracy of the titration was compared with that of acidimetric coulometric titration of ammonia.
More Related Videos
Related Concept Videos
Titration of a Weak Base with a Strong Acid
Using the ICE table and substituting the Kb value, we calculate the initial pH of 50 mL of 0.1 M ammonia to be 11.11. Addition of 25 mL of 0.1 M hydrochloric acid to this solution of ammonia results in a buffer with an equal concentration of ammonia and ammonium ions. The pH of this buffer can be calculated by substituting these...