Tuning a Parallel Segmented Flow Column and Enabling Multiplexed Detection
Published on: December 15, 2015
Cross-correlation in flow-injection analysis with parallel flow streams and amperometric detection
1Department of Chemistry, University of Massachusetts, Amherst, MA 01003, U.S.A.
Abstract:
Cross-correlation was implemented for flow-injection analysis by using two parallel flow lines, each with amperometric detectors, and driven by peristaltic pumps. One flow line was used to generate the reference signal for an analog correlator circuit and the other to generate the analyte signal. Cross-correlation was performed by multiplying these signals together at a time delay of zero, followed by low pass filtering. Using dopamine as a test system, improvements in signal-to-noise ratios of about two orders of magnitude were found for the correlation signal over the direct measurement of the electrode current.