Application of the slotted quartz tube in flow-injection flame atomic-absorption spectrometry
1Flow-Injection Analysis Research Centre, Institute of Applied Ecology, Academia Sinica, Box 417, 110015 Shenyang, China.
Abstract:
The slotted quartz has been applied to flow-injection flame atomic-absorption spectrometry (FI-FAAS) showing several important advantages. The tube life was improved by a factor of 5-6 compared to conventional continuous aspiration. Flow impact systems were found not to be necessary in the applications so that larger enhancement factors may be achieved without sacrifice in precision. For 1.0 mg/l. copper, 1.0 mg/l. lead, 0.1 mg/l. cadmium and 1.0 mg/l. gold sensitivity enhancement factors of 3.1, 5.5, 5.3 and 4.0 were obtained with precisions of 1.3%, 1.1%, 1.6% and 1.7% RSD (n = 11) respectively. Application of the slotted quartz tube FI-FAAS method to the determination of heavy metals in urine has shown improved tolerance to interfering matrices. Recoveries obtained by spiking undiluted urine samples with 0.1 mg/l. copper and lead, and 0.01 mg/l. cadmium were in the range 100-102%.
Related Concept Videos
Atomic Fluorescence Spectroscopy
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Emission Spectroscopy: Instrumentation
Flame Photometry: Overview