Related Experiment Video
Updated: Jun 28, 2026

Multi-analyte Biochip (MAB) Based on All-solid-state Ion-selective Electrodes (ASSISE) for Physiological Research
Published on: April 18, 2013
A double-membrane ephedrine selective electrode based on ephedrine-tetraphenylborate in poly (vinyl chloride) resin
1Department of Experimental Didactic Sciences, University of Zaragoza, Huesca, Spain.
Abstract:
The construction and performance characteristics of a double-membrane ephedrine selective electrode are described. The electrode is based on the use of an internal conducting membrane made of tetrabutylammonium bromide and another external electroactive membrane containing the ionic pair ephedrine-tetraphenylborate in a poly (vinyl chloride) resin (with some plasticizers incorporated) as an inert matrix. The ephedrine electrode exhibits linear Nernstian response within the range 10(-2)-5 x 10(-5)M of ephedrine, with a slope of 58.2 +/- 0.5 mV decade(-1) at 25 +/- 0.2 degrees . The detection limit is 10(-4.5)M. The reproducibility (coefficient of variation) was 0.54% (n = 10 determinations) and the stability of its potential is 2.3 mV/24 h. The selectivity coefficients for 15 ions were calculated. The electrode was applied to the determination of ephedrine in some pharmaceutical preparations with satisfactory results.
Related Concept Videos
Potentiometry: Membrane Electrodes
Ion-Exchange Chromatography
The Electrical Double Layer
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrodeposition
Electrodeposition can...
Ion Exchange