Related Experiment Video
Updated: Jun 28, 2026

Simple Methods for the Preparation of Non-noble Metal Bulk-electrodes for Electrocatalytic Applications
Published on: June 21, 2017
Preparation and characterization of a new enzyme electrode based on solid paraffin and activated graphite particles
C Petit1, A Gonzalez-Cortes, J M Kauffmann
1Free University of Brussels, Institute of Pharmacy, Bd. Triomphe, Campus Plaine, CP 205/06, 1050 Brussels, Belgium.
Abstract:
A new electrochemical biosensor was developed by incorporating an enzyme into a solid-paraffin-graphite-particle matrix. Tyrosinase served as model enzyme and the biosensor response was characterized with respect to its response to dopamine. The influence of different experimental parameters (tyrosinase loading, flow rate, oxygen dependence, pH, etc.) was investigated in order to optimize the biosensor performance. The electrode response was fast, reversible and linear in a large concentration domain (0.1 muM-1 mM). The enzyme-solid paraffin carbon paste electrode (CPE) showed markedly improved stability in flow injection analysis compared to the classical liquid paraffin-graphite-based biosensors. The biosensor allowed a sampling rate of 79 samples per hour, the repeatability of the injections was improved with respect to the classical CPE with a relative standard deviation of 2.2% (N = 63), and the detection limit for dopamine was 50 nM. The biosensor response to some phenol and catechol derivatives was also investigated.