Related Experiment Video
Updated: Jun 28, 2026

Characterization of Nanocrystal Size Distribution using Raman Spectroscopy with a Multi-particle Phonon Confinement Model
Published on: August 22, 2015
Quantitative determination of silicon in silica dust by FT-Raman spectroscopy
Abstract:
Silicon was quantitatively determined in respirable silica dusts and silica fume by FT-Raman spectroscopy using barium sulfate as an internal standard. Barium sulfate was selected as the standard as its Raman spectrum does not overlap with that of silicon and its scattering cross-section is significantly lower. This allowed silicon admixed with barium sulfate in ratios as low as 0.00003 to be detected. The laser beam was defocussed to cover the sample area. 128 scans at 4 cm(-1) resolution with 600 mW of power at the sample achieved a detection limit of 0.1 mug of silicon in a 5 mg sample. Silicon to BaSO(4) ratios in the range 0.00003-0.06 showed a near linear response with any decade change being linear. The RSD for this analytical method was +/-2%. Silica fume and respirable silica dust showed silicon concentrations to be in the range 2.6-51.5 ppm and 0.05-0.7 mg m(-3) respectively.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...